GOG-L oxygen measuring set for low oxygen concentrations

Article no.
gr-605926
€1,070.17 €899.30
Delivery time:
2-4 days
Complete set of measuring device GMH 3692, sensor, hand pump, puncture needles and case. For measuring the residual oxygen content in packaging.
GOG-L oxygen measuring set for low oxygen concentrations is available to buy in increments of 1

Complete set for determining the residual oxygen content in packaging

The hand-held meter GMH 3692 reliably determines the residual oxygen content in packaging. It consists of the air oxygen meter GMH 3692, a hand pump and an oxygen sensor with piercing needle, installed in an external sensor housing.

Features hand-held meter GMH 3692:

  • Double display for oxygen and temperature

  • Measured variables: O 2 concentration and O 2 partial pressure

  • Alarm monitoring with integrated horn

  • Automatic temperature compensation

  • Min / Max value memory, hold function

  • serial interface, can be connected directly to the RS232 or USB interface of a PC via a galvanically isolated interface converter (optionally available, see accessories)

  • Bus-capable device: up to 5 devices can be connected to a PC interface

  • Battery and power supply operation possible

  • Operating elements: 6 membrane keys for on / off switch, selection of measuring range, min. And max. Value memory, hold function, calibration, etc.

  • Calibration: To check the accuracy, we optionally offer calibration bags

    • 1-point calibration: simple quick calibration to atmospheric air (device is adjusted to 20.9% by pressing a button).
    • 2-point calibration: first point in air (20.9%), second point freely selectable.

Features oxygen sensor GGO 381:

  • closed sensor version

  • built into housing, replaceable

  • Lifetime: 12 months warranty on sensor element (if used properly and ambient pressure)

  • Device connection via an approx. 1.3 m long cable with mini DIN plug

applications

  • Biochemistry: Oxygen monitoring in incubators for the cultivation of cell cultures. Monitoring the fermentation of fruits in fermentation plants, etc.

  • Medical technology: monitoring of the oxygen concentration in ventilators; Breath control; Oxygen concentration monitoring eg in oxygen tents, etc.

  • Food technology: Checking the residual oxygen in packaging (eg coffee, tea, MAP, etc.). Monitoring the oxygen content of critical production processes and storage.

  • Air conditioning and ventilation technology: oxygen measurement; Air quality monitoring; Oxygen concentration measurement in closed ventilation systems, etc.

  • Sport: Immersion gas test: checking the oxygen concentration in nitrox, trimix or similar gas mixtures

Note:

Not suitable for use in "underwater applications" (rebreather etc.) Air oxygen meter without sensor

Specifications
Use Hand-held devices
Measured Variable Air oxygen
Operating conditions Temperature: 0 to +50 ° C
Relative humidity: 0 to 95% RH (non-condensing)
Accuracy O2 concentration: ± 0.1% ± 1 digit
Temperature: ± 0.1 ° C ± 1 digit
Air pressure: ± 3 hPa or 0.1% of MW (higher applicable)
Measuring range O2 concentration: 0.0 ... 95.0 % O2 (gaseous), 0 ... 1100 hPa O2
Temperature: -5.0 ... +50.0
Air pressure: GMH 3692: 10 ... 1200 hPa
GMH 3695: 300 ... 5000 hPa
Power supply 9 V battery as well as additional d.c. connector for external 10.5 ... 12 V direct voltage supply. (suitable power supply:GNG10/3000)
Battery Life approx. 300 h
Control Elements 6 membrane switches for power on/off, range selection, min and max value storage, hold function, calibration, etc.
Data transfer USB cable, Serial Interface
Dimensions Unit: 142 x 71 x 26 mm Softcase: 394 x 294 x 106 mm
Case ABS
Weight ca. 160 g (incl. battery)