G 1690 Air Oxygen Meter (former GOX100)
Article no.
GR-482286
€244.07
€205.10
Delivery time:
1-10 days
Compact handheld meter for O2 incl. Sensor, hose adapter and T-piece
G 1690 Air Oxygen Meter (former GOX100) is available to buy in increments of 1
Handheld measuring device for atmospheric oxygen
The G 1690 is suitable for universal use, but not for use in "underwater applications" (rebreathers, etc.).
- Min / Max Value Memory
- Sensor element interchangeable without tools
- Easy calibration in the ambient air
- Sensor connection via a connection cable with a jack plug, which is approx. 0.7 m long and permanently connected to the device
- Battery change indicator "BAT"
Note: Before the measurement and calibration, the protective cap of the oxygen sensor must be removed! Screwing on the cap for storage increases sensor life.
Delivery contents:
Handheld device with sensor GOEL 370 (not included in -GL version)
Flow-Diverter ESA 369, T-piece ZOT 369
2xAA batteries
Operating manual
Additional accessories depending on the model
Note: The sensor should be exchanged every 2 years or if the sensor capacity is below 70% (see manual).
Use | Hand-held devices |
---|---|
Measured Variable | Air oxygen |
Operating conditions | Temperature: 0 ... 45 °C (Sensor); recommended 5..30 °C Relative humidity: 0 ... 95 % RH Over- / under pressure: ±0,25 bar (pressure difference) Ambient pressure: 0,6 ... 1,75 bar abs Storing temperature: -20 ... +50 °C |
Accuracy | ± 0.1% O2 ± 1 digit, with calibrated device (in the range 15 ... 40% O2) Sensor GOEL 370 <2 % O2 ±0,2 % O2 <25 % O2 ±0,5 % O2 >25 % O2 ±0,5 % O2 |
Resolution | 0,1 Vol.-% O₂ |
Measuring range | 0,0 ... 100,0 Vol.-% O2, recommended for 0,2 ... 35,0 % Vol.-% O₂ Oxygen partial pressure: 0 ... 1100 hPa O2 Recommended oxygen concentration: 0.2 ... 35% O2 |
Measuring interval | ca. 1 measurement per second |
Degree of protection | IP20, IP30, IP40 |
Sensors | Electrochemical oxygen partial pressure sensor in external sensor housing, M16x1 connection thread, connection 0.7 m cable with jack plug |
Initialisation time | 90% in < 10 sec., depends on temperature |
Power supply | 2x AA battery (incl) |
Battery Life | >3000 h |
Data transfer | None |
Dimensions | 108 x 54 x 28 mm (H x W x D) without Sensor |
Case | ABS |
Weight | 175 g |